Model | DP6350B | ||
Bandwidth(-3dB) | 300MHz | ||
Rise time | ≤1.16ns | ||
Accuracy | ±2% | ||
Range selection (Attenuation rate) | 100X/1000X | ||
Maximum differential voltage(DC + Peak AC) | 100X | ±350V | |
1000X | ±3500V | ||
Common mode voltage(DC + Peak AC) | ±3500V | ||
Maximum differential mode VS frequency curve | Figure 3 | ||
Maximum input voltage-to-earth (Vrms) | 600V CATIII 1000V CATII | ||
Input impedance | Single-ended to ground | 5MΩ | |
Between inputs | 10MΩ | ||
Input capacitance | Single-ended to ground | <4pF | |
Between inputs | <2pF | ||
CMRR | DC | >80dB | |
100kHz | >60dB | ||
1MHz | >50dB | ||
Noise(Vrms) | 100X | <350mV | |
1000X | <550mV | ||
Differential overvoltage detection level | 100X | ≥350V | |
1000X | ≥3500V | ||
Propagation time | Probe | About 6ns | |
BNC Line(1m) | About 5ns | ||
Bandwidth limit filters (5MHz) | ≥-3dB@5MHz | ||
Overload indicator (red light) | Yes | ||
Overload alarm | Yes(Can shut up manually) | ||
Automatic save | Yes | ||
Offset setting function | Yes( Set in test mode) | ||
Terminate load | ≥100kΩ | ||
Power supply | USB 5V/2A adapter | ||
Safety standard | EN61010-1: 2010 | ||
EMC standard | EN61326-1:2013 EN61000-3-2:2006+A1:2009+A2:2009 EN61000-3-3:2013 |