DP6070 (700V/70MHz)
DP6070 (700V/70MHz)

Rise Time ≤5ns

Accuracy ±2%


浏览次数:0
说明书: 产品说明书

1. Summary

DP6000 Series High Voltage Differential Probes are designed for the measurement of high voltage differential signal, to meet the demand for floating measurement. 

The probes have a max 200MHz bandwidth and 7000V differential voltage which can meet the majority measurement applications. 

The users can adjust the offset output voltage of the probe.

The 5MHz bandwidth limit function can remove the higher frequency noise and interference; 

The probe has sound & light alarming function which can also be turned off manually 

It can be powered by USB port, which is convenient for the customers;

The probe is standard BNC output port which can be connected to oscilloscope of any brand. 

The probes have great CMRR which is very important in many power electronics applications.


1、Electric Specification

ModalDP6070
Bandwidth(-3dB)70MHz
Rise time≤5ns
Accuracy±2%
Range selection(Attenuation rate)10X/100X
Maximum differential voltage(DC + Peak AC)10X±70V
100X±700V
Common mode voltage(DC + Peak AC)±700V
Maximum differential mode VS frequency curveReference Figure 1
Maximum input
voltage-to-earth (Vrms)
450V CATII
600V CATI
Input impedanceSingle-ended to ground

2.5MΩ

Between inputs

5MΩ

Input capacitanceSingle-ended to ground<4pF
Between inputs<2pF
CMRRDC>80dB
100kHz>60dB
1MHz>50dB
Noise(Vrms)10X<20mV
100X<120mV
Differential overvoltage detection level10X

70V

100X≥700V
Propagation  timeProbeAbout 9ns
BNC Line(1m)About 5ns
Bandwidth limit filters (5MHZ)

≥-3dB@5MHz

Overload indicator (red light)Yes
Overload alarmYes(Can shut up manually)
Automatic saveYes
Offset setting functionYes( Set in test mode)
Terminate load

≥100kΩ

Power supplyUSB 5V/1A adapter
Safety standardEN61010-1: 2010
EMC standardEN61326-1:2013  EN61000-3-2:2006+A1:2009+A2:2009  EN61000-3-3:2013